Starten Sie Ihre Suche...


Durch die Nutzung unserer Webseite erklären Sie sich damit einverstanden, dass wir Cookies verwenden. Weitere Informationen

Hochaufgelöste Phasenkontrastmessungen zur Ermittlung der Druckverhältnisse im kleinen Kreislauf bei Patienten mit chronisch thromboembolischer pulmonaler Hypertonie – Korrelation mit simultan ermittelten, invasiven Druckwerten

Laufzeit: 01.01.2006 - 31.12.2009

Kurzfassung


Purpose: to estimate mean pulmonary arterial pressure (mPAP) in patients with CTEPH by analysing flow parameters of the pulmonary trunc derived from high temporal resolution phase-contrast MR-Imaging (PC-MRI).
Materials and Methods: We prospectively examined 8 CTEPH patients with high temporal resolution PC-MRI at 1.5 T (Magnetom Sonata , Siemens Medical Solutions) with simultaneous invasive pressure recordings in the pulmonary trunc. Imaging parameters for PC-MRI were as follows: TR = 10...
Purpose: to estimate mean pulmonary arterial pressure (mPAP) in patients with CTEPH by analysing flow parameters of the pulmonary trunc derived from high temporal resolution phase-contrast MR-Imaging (PC-MRI).
Materials and Methods: We prospectively examined 8 CTEPH patients with high temporal resolution PC-MRI at 1.5 T (Magnetom Sonata ®, Siemens Medical Solutions) with simultaneous invasive pressure recordings in the pulmonary trunc. Imaging parameters for PC-MRI were as follows: TR = 10 msec, TE = 2.5 msec, Flip = 15 °, velocity encoding = 100 cm/sec, bandwidth = 977 Hz/pixel, 3 averages. The parameters absolute and relative acceleration time ATa [mesc] and ATr [%], peak of mean velocities MRV [cm/sec], and distensibility DIS [%] were evaluated according to Proc. Intl Soc Mag Reson Med 2003; 11: 407, and were used to compute mPAP with the following regression equation: mPAP–PC-MRI [mmHg] = 88.9- 0.29 * ATa – 1.1 * ATr – 0.38 * MFV – 0.16 * DIS.
Results: MPAP-MRI correlated very well with simultaneous invasive pressure recordings in all patients (linear coefficient of determination R2 = 0.979, p < 0.001, median difference +2.0 mm Hg, range –1.9 - + 4.7 mm Hg).
Conclusion: Non-invasive mPAP-estimations are possible in CTEPH-patients by means of high temporal resolution PC-MRI and correlate well with simultaneous invasively measured values.
» weiterlesen» einklappen

Veröffentlichungen


Beteiligte Einrichtungen